XARTU/5™ Manual Copyright© 2013
Eagle Research Corporation
®
REV 11.1.11 Page | 34
3. The unit will enter calibration mode and the display will show the current value and
mA representation of the analog output signal. For example,
Eng: 48.000
mA: 11.680
where 48.000 is the analog output reading representing 11.680 mA. The top line
will alternate between four different readouts Eng: 48.000, UP/DN TO
CHANGE,
CALIBRATING, and the parameter label while the bottom line will always show
the mA value.
4. Connect a multimeter in series with the loop to measure the current. The field
instrument that the loop is driving can also be used to read the output, if desired.
5. Pressing will increment the output current to represent 0%, 25%, 50%, 75%, &
100% of the analog output parameter to check the calibration. Pressing will
decrement the output current. Therefore, 0% = 4mA, 25% = 8mA, 50% = 12mA,
75% = 16mA, & 100% = 20mA.
6. If adjustments are needed, press zero. The display now shows:
zero: x.xxx
04.000 +y.yyy mA
x.xxx represents the default zero value (low scale), and y.yyy is the adjustment
made to 4mA for the analog output signal. The adjustment can either be positive
or negative shown by + or - respectively. The top line will alternate between
zero: x.xxx and UP/DN TO ADJUST. This is the zero adjusted value to calibrate
the analog output to 4mA.
7. Press or to increase or decrease the output until the meter reads 4mA, or the
current zero reference matches the field equipment. Press ent to collect the new
point.
8. Press span. The display shows:
span: x.xxx
20.000 +y.yyy mA
where x.xxx represents the default span value (full scale), and y.yyy is the
adjustment made to 20mA for the analog output signal. The adjustment can
either be positive or negative shown by + or - respectively. The top line will
alternate between span: x.xxx and UP/DN TO ADJUST. This is the span
adjusted value to calibrate the analog output to 20mA.